Technical data

566 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
27 Using the STATus System
:STATus:QUEStionable:CALibration:PTRansition?
Determines which bits in the Questionable Calibration Condition register will set the
corresponding bit in the Questionable Calibration Event register when the condition register
bit has a positive transition (0 to 1). The variable <number> is the sum of the decimal values
of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Calibration Extended Needed Register
Questionable Calibration Extended Needed Condition
:STATus:QUEStionable:CALibration:EXTended:NEEDed
:CONDition?
Returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Needed Condition register.
The data in this register is continuously updated and reflects the current conditions.
Key Type: There is no equivalent front-panel key.
Questionable Calibration Extended Needed Enable
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle
<number>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Determines which bits in the Questionable Calibration Extended Needed Enable Register
will set bits in the Questionable Calibration Event register, which also sets the Calibration
Summary bit (bit 8) in the Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type: There is no equivalent front-panel key.
Example STAT:QUES:CAL:EXT:NEED:ENABLE 16384 could be used if you
have turned off the automatic alignment and you want to query whether an alignment is
needed.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767