Technical data

Using the STATus System 27
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 565
:STATus:QUEStionable:CALibration:ENABle?
Determines which bits in the Questionable Calibration Condition Register will set bits in the
Questionable Calibration Event register, which also sets the Calibration Summary bit (bit 8)
in the Questionable Register. The variable <number> is the sum of the decimal values of the
bits you want to enable.
Key Type: There is no equivalent front-panel key.
Example STAT:QUES:CAL:ENABLE 16384 could be used if you have turned off
the automatic alignment and you want to query whether an alignment is needed.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Calibration Event Query
:STATus:QUEStionable:CALibration[:EVENt]?
Returns the decimal value of the sum of the bits in the Questionable Calibration Event
register.
The register requires that the associated PTR or NTR filters be set before a condition
register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the register is cleared.
Key Type: There is no equivalent front-panel key.
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition <number>
:STATus:QUEStionable:CALibration:NTRansition?
Determines which bits in the Questionable Calibration Condition register will set the
corresponding bit in the Questionable Calibration Event register when the condition register
bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type: There is no equivalent front-panel key.
Factory Preset: 0
Range: 0 to 32767
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition <number>