Technical data

Using the STATus System 27
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 557
*OPC?
Sets bit 0 in the standard event status register to “1” when pending operations have
finished. It does not hold off subsequent operations.
The *OPC? query stops new commands from being processed until the current processing
is complete. Then it returns a “1”, and the program continues. This query can be used to
synchronize events of other instruments on the external bus.
The instrument does not wait for completion of all processes for these commands. The
processes that are monitored are identified in the STATus:OPERation register. These
include:
For example, if you want to verify the completion of both calibration and waiting for trigger
set :STAT:OPER:ENAB 33 and monitor any changes.
Key Type: There is no equivalent front-panel key.
Query Instrument Options
*OPT?
This command is not implemented. However the information is available remotely by using
SCPI commands. See :SYSTem:OPTions? in “Show System" on page 494.
Power-On Status Clear
*PSC
*PSC?
Sets the state of the power-on status clear flag. This command allows you to specify if the
service request enable register and the event status enable register should be cleared at
power up.
The query returns the contents of the power-on status clear flag.
Key Type: There is no equivalent front-panel key.
Process STATus:OPER
Register Bit
Byte Value
Calibrating 0 1
Sweeping 3 8
MEASuring (not in all modes) 4 16
Waiting for trigger 5 32