Technical data
Trig 26
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 535
Trig Dela y
Allows you to control a time delay during which the N8201A will wait to begin a sweep after
receiving an external or line trigger signal. You can use negative delay to pre-trigger the
instrument.
Trigger Delay is not available in Free Run, so turning Free Run on turns off Trigger Delay, but
preserves the value of Trigger Delay.
Key Path: Tr ig
Dependencies/Couplings:
• This function is not available when Trigger is Free Run or Video.
• This function is not available when Gate is on.
State Saved: Saved in Instrument State.
Factory Preset: Off, 1 μs
Range: –150 ms to +500 ms
Remote Command:
:TRIGger[:SEQuence]:DELay <time>
:TRIGger[:SEQuence]:DELay?
:TRIGger[:SEQuence]:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:DELay:STATe?
Example:
TRIG:DEL:STAT ON
TRIG:DEL 100 ms