Technical data
520 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
25 Trace/View
Operations
This menu allows the user to do simple trace operations.
Key Path: Trace/View
1 <-> 2
Exchanges the contents of trace 1 with the contents of trace 2 and puts both in view mode.
The action is performed once. It is not available for continuous sweeps. To clear any trace
operation, set the trace back to Clear Write, Max Hold, or Min Hold.
The X-Axis settings and domain of a trace stay with it when it is copied or exchanged.
Key Path: Trace/View, Operations
Remote Command:
:TRACe:EXCHange: <trace_name>, <trace_name>
Example:
:TRAC:EXCH TRACE1,TRACE2
2 – DL -> 2
Subtracts the display line from trace 2 and places the result in trace 2. The action is
performed once. It is not available for continuous sweeps. To keep the modified trace 2 from
changing after the operation, it is set to View. To clear any trace operation, set the trace
back to Clear Write, Max Hold, or Min Hold.
Key Path: Trace/View, Operations
Remote Command:
:TRACe:MATH:SUBTract:DLIN <trace_name>
Example:
TRAC:MATH:SUBT:DLIN TRACE2
2 <-> 3
Exchanges the contents of trace 2 with the contents of trace 3 and puts both in view mode.
The action is performed once. It is not available for continuous sweeps. To keep the
modified traces from changing after the operation, they are set to View. To clear any trace
operation, set the trace back to Clear Write, Max Hold, or Min Hold.