Technical data
Amplitude / Y Scale 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 49
Scale/Div
Sets the units per vertical graticule division on the display. This function is only available
when Scale Type (Log) is selected and the vertical scale is power, or Demod View is on and
the vertical scale is hertz. When Scale Type (Lin) is selected, Scale/Div is unavailable.
Key Path: AMPLITUDE / Y Scale
State Saved: Saved in Instrument State
Factory Preset: 10 dB, in logarithmic scale type
Fundamental Units: dB; kHz
Te rm in a tor s: dB; Hz, kHz, MHz, GHz
Default Terminator: dB; kHz
Range: 0.1 dB to 20 dB, 1 kHz to 240 kHz (When Demod View (On) is selected.)
Remote Command:
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision <power>
:DISPlay:WINDow[1]:TRACe:Y:[SCALe]:PDIVision?
Example:
DISP:WIND:TRAC:Y:PDIV 5 DB