Technical data

486 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
24 System
Frequency Corrections (Remote Command Only)
Turns the internal frequency corrections on/off.
Saved State: Not saved in instrument state
Factory Preset: On
Remote Command:
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Example:
CAL:FREQ OFF
Align Subsys
Accesses the keys to immediately execute an alignment of one of the subsystems (Align RF,
Align IF, Align ADC, and Align Current Sys Gain). When one of the subsystem alignments is
started, all other operations are stopped and any alignment messages will be visible on the
display.
Key Path: System, Alignments
Remote Command:
There is no remote command for this key.
Align RF
Initiates an alignment on the RF assembly.
Key Path: System, Alignments, Align Subsys
Annunciation/Annotation: Displays a series of pop-up message boxes, indicating an
alignment is in progress.
If a signal is present that interferes with the alignment a message Align RF 50 MHz
signal detected. will be displayed.
SCPI Status Bits/OPC Dependencies: Status questionable calibration. Bits 3, 11, or 12
will be set if
Align RF fails
Remote Command: :CALibration:RF
:CALibration:RF?