Technical data
System 24
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 479
Factory Default: Mode
Remote Command: Preset Type
Example:
SYST:PRES[:USER]:SAVE saves the current state to be used as the preset user state.
SYST:PRES:TYPE USER defines the type of preset as the user preset.
With user preset selected, and a user state saved, use SYST:PRES to do a user preset.
Mode
Sets the preset type to “Mode.” When you do a preset, the current mode factory default
instrument state will be restored. A mode preset does not change the mode. Refer to
“Preset" on page 441 for more information.
Key Path: System, Power On/Preset, Preset Type
Readback: Mode
Dependencies/Couplings: None
State Saved: Survives Preset and power cycle, but not saved in Instrument State.
Factory Default: Mode
Remote Command: :SYSTem:PRESet:TYPE “Preset Type”
Example:
SYST:PRES:TYPE MODE defines the type of preset as the mode preset.
After you have selected mode as the preset type, use SYST:PRES to do a mode preset.
Factory
Sets the preset type to “Factory.” When you do a preset, all of the factory default
instrument state will be restored. A factory preset switches the N8201A to the Spectrum
Analysis mode and resets the settings of all the modes to the factory defaults (that is,
Spectrum Analysis Mode with continuous sweep). Refer to “Preset" on page 441 for more
information.
Key Path: System, Power On/Preset, Preset Type
Annunciation/Annotation: Factory
State Saved: Survives Preset and power cycle, but not saved in Instrument State.