Technical data

478 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
24 System
Example:
SYST:PON:TYPE LAST defines the power on type as the last state the N8201A was in
before power was turned off.
SYST:PON:TYPE?
Preset Type
Enables you to select what type of preset will be initiated when you press the green Preset
key or send the remote command, using SYST:PRES. Choose between Factory, Mode,
or User defined presets.
Key Path: System, Power On/Preset
State Saved: Survives Preset and power cycle, but is not saved in Instrument State.
Preset: Does not change the Preset Type.
Factory Default: Mode
Remote Command:
:SYSTem:PRESet:TYPE FACTory|USER|MODE
:SYSTem:PRESet:TYPE?
Remote Command Notes:
SYSTem:PRESet:USER:SAVE defines the user preset.
Example:
SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
User
Sets the preset type to User. When Preset is pressed, the instrument state that you have
defined as the user preset setting is restored. Use the Save User Preset key to define your
user preset settings. Refer to “Preset" on page 441 for more information. If you have not
saved a user state, then the instrument will save the power-up state for you to use as a
default user preset state.
Key Path: System, Power On/Preset, Preset Type
Readback: User
State Saved: Survives Preset and power cycle, but not saved in Instrument State.