Technical data

468 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
23 Sweep
Key Path: Sweep, Gate Setup
State Saved: Saved in Instrument State
Remote Command:
[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0
[:SENSe]:SWEep:EGATe:VIEW?
Example:
SWE:EGAT:VIEW ON
Polarity
Sets the polarity for the gate signal. When Positive (Pos) is selected, a positive-going edge
will satisfy the gate condition, after the delay is set with the Delay key. When
Negative (Neg) is selected, a negative-going edge will satisfy the gate condition after the
delay.
Key Path: Sweep, Gate Setup
State Saved: Saved in Instrument State
Factory Preset: Positive
Remote Command:
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive?
Example:
SWE:EGAT:POL NEG
SWE:EGAT:POL?
Delay
Controls the length of time from the time the gate condition is satisfied until the gate is
enabled.
Key Path: Sweep, Gate Setup
State Saved: Saved in Instrument State
Factory Preset: 57.7 µs