Technical data

424 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
17 Mode Setup (Spectrum Analysis Mode)
Retain Params
Enables you to retain the base-instrument parameter settings used by individual
measurements that are typically lost when switching between measurements.
When
Retain Params is set to Off, the values of the fully configurable parameters are not
saved and recalled during a measurement change.
When
Retain Params is set to On, the fully configurable parameters will be saved and
recalled during a measurement change, along with
Span, Ref Level, Single/Cont,
Detector, Avg/VBW Type, Points, and if applicable, Spectrum/Bar/Combined and
Graticule On/Off.
Key Path: Mode Setup
Factory Preset: Off
Remote Command:
[:SENSe]:RADio:STANdard:RPARams[:STATe]ON|Off|1|0
[:SENSe]:RADio:STANdard:RPARams[:STATe]?
Example:
RAD:STAN:RPAR Off
NOTE
When Radio Std or Radio Std Setup is changed, Retain Params is set to Off. Once
Retain Params has been set to Off, parameters will no longer be retained and any settings
you may have entered will be lost.