Technical data
414 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
17 Mode Setup (Spectrum Analysis Mode)
GSM/EDGE
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
GSM/EDGE. Power Stat CCDF and Burst Power are the only measurements available for
this standard.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN GSM
3GPP W-CDMA
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
W-CDMA. All measurements except Harmonic Distortion and Spurious Emissions are
available for this standard.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN W3GPP
cdma2000 MC-1X
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
cdma2000-MC-1X. All measurements except Harmonic Distortion, Spurious Emissions, and
Spectrum Emission Mask are available for this standard.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std