Technical data

412 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
17 Mode Setup (Spectrum Analysis Mode)
Radio Std
Accesses the available radio standard. Selecting a radio standard modifies spectrum
analyzer settings for the measurements activated under the
MEASURE menu.
Measurements that are not relevant to the selected radio standard are not available (grayed
out) and they do not have instrument default settings associated with them. If
Enable All
Measurements
is set to Yes when the Radio Std is pressed, only the standards that are
relevant to the currently active measurement will be available. Unavailable standards will
be grayed out.
Key Path: Mode Setup
Factory Preset: None
Remote Command:
[:SENSe]:RADio:STANdard[:SELect]
NONE|IS95a|JSTD|IS95C|GSM|W3GPP
|C2000MC1|C2000DS|C2000MC3|NADC|PDC|BLUEtooth|TETRa|WL802D
OT11A
|WL802DOT11B|WL802DOT11G|WL802DOT11N|HIPERLAN2|DVBTDLN
DVBTGPN|DVBTIPN|FCC15|SDMBSE|UWBINDOOR|
None
Selects no radio standard. If Radio Std, None is selected when a measurement is running,
all measurement parameters are returned to their default values. Analyzer parameters
outside the
MEASURE or Meas Setup key menus are not affected.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN NONE
IS-95A
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
IS-95A. The Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions
measurements are not available when this radio standard is selected.