Technical data

398 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Abs OR Rel - Fail is shown if one of the absolute spectrum emission mask measurement
results is larger than the limit between Abs Start and Abs Stop, inclusive OR one of the
relative spectrum emission mask measurement results is larger than the limit for Rel
Start and Rel Stop.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST
ABSolute|AND|OR|RELative
[:SENSe]:SEMask:OFFSet[n]:LIST:TEST?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.
Example:
SEM:OFFS:LIST:TEST ABS,REL,AND,OR
SEM:OFFS:LIST:TEST?
Absolute
Set the limit test to show Fail if one of the absolute spectrum emission mask measurement
results is larger than the limit for Abs Start and/or Abs Stop. This is the default selection for
each offset
Key Path: Meas Setup, Offset/Limits, Fail Mask
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Remote Command: Refer to “Fail Mask" on page 397.
Relative
Sets the limit test to show Fail if one of the relative spectrum emission mask measurement
results is larger than the limit for Rel Start and/or Rel Stop.
Key Path: Meas Setup, Offset/Limits, Fail Mask