Technical data

392 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.
Example:
SEM:OFFS:LIST:SWE:AUTO 4 ms
SEM:OFFS:LIST:SWE:AUTO?
SEM:OFFS:LIST:SWE 4 ms
SEM:OFFS:LIST:SWE?
Res BW
Specifies the resolution bandwidth used in measuring the offset pair. When set to Auto, the
Res BW from the default tables for “Offset/Limits" on page 388 are used. When set to
Man, the range of settings is the range of available Res BWs of the N8201A, except the
maximum is further limited to not exceed (Stop Freq - Start Freq).
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Default Terminator: Hz
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution] <freq>...[<freq>] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution]?
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution]:AUTO?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.