Technical data
390 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Start Freq
Specifies the inner limit (frequency closest to the carrier) for both segments of the specified
offset pair. When Start Freq (Off) is selected, the offset pair is not measured.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Default Terminator: Hz
Range: 0 Hz to 500 MHz
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt <freq>, ...
[<freq>] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:FREQuency:STARt?
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe OFF|ON|0|
[:SENSe]:SEMask:OFFSet[n]:LIST:STATe?
Remote Command Notes:
Comma separated list. When offset F is off, this will be 5 values. When offset F is on,
this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is BTS.
Offsets that are turned off, for the currently selected offset will return a
– 999.0 when queried.
Example:
SEM:OFFS:LIST:FREQ:STAR 2 MHz
SEM:OFFS:LIST:FREQ:STAR?
SEM:OFFS:LIST:STAT 1
SEM:OFFS:LIST:STAT?
Stop Freq
Specifies the outer limit (frequency furthest from the carrier) for both segments of the
specified offset pair. The lower range is limited to the setting of Start Freq.
Key Path: Meas Setup, Offset/Limits