Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 389
Offset
Selects the offset pairs (upper and lower) that the menu keys affect, and displays the
memory selection menu from A to F for (where you can store up to 6 sets of values for Start
Freq, Stop Freq, Sweep Time, Res BW, Meas BW, Abs Start, Abs Stop, and so forth). Press
Offset until the letter of the desired offset (A, B, C, D, E, or F) is underlined. Only one
selection at a time is shown on this key label. Since the display can show only five results at
one time, the Results Index key allows you to scroll to the sixth result when all offsets are
active.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: A
Remote Command: There is no remote command for this function.
C, On 20.0 30.0 100 100 − 12.0 − 24.0 Man - 28.0 − 40.0 Man Rel 1
D, On 30.0 50.0 100 100 − 24.0 − 24.0 Couple − 40.0 − 40.0 Couple Rel 1
E, Off 50.0 100.0 100 100 − 24.0 − 24.0 Couple − 40.0 − 40.0 Couple Rel 1
F, Off 216.0 250.0 100 100 − 24.0 − 24.0 Couple − 47.0 − 47.0 Couple Rel 1
Table 10 Offsets & Limits Defaulted for 802.11a, 802.11g, and HiperLAN/2 Radio Standards
Offset Start
Freq
(MHz)
Stop
Freq
(MHz)
Sweep
Time
(ms)
Res BW
(kHz)
Abs
Start
(dBm)
Abs
Stop
(dBm)
Abs
Couple
Rel
Start
(dBc)
Rel
Stop
(dBc)
Rel
Couple
Fail
Mask
Meas
BW
Table 11 Offsets & Limits Defaulted for 802.11b
Offset Start
Freq
(MHz)
Stop
Freq
(MHz)
Sweep
Time
(ms)
Res BW
(kHz)
Abs
Start
(dBm)
Abs
Stop
(dBm)
Abs
Couple
Rel
Start
(dBc)
Rel
Stop
(dBc)
Rel
Couple
Fail
Mask
A, On 11.0 22.0 100 100 − 10.0 − 10.0 Couple − 30.0 − 30.0 Couple Rel
B, On 22.0 50.0 100 100 − 30.0 − 30.0 Couple − 50.0 − 50.0 Couple Rel
C, On 50.0 70.0 100 100 − 30.0 − 30.0 Couple - 50.0 − 50.0 Couple Rel
D, On 70.0 90.0 100 100 − 30.0 − 30.0 Couple − 50.0 − 50.0 Couple Rel
E, Off 90.0 100.0 100 100 − 30.0 − 30.0 Couple − 50.0 − 50.0 Couple Rel
F, Off 100.0 120.0 100 100 − 30.0 − 30.0 Couple − 50.0 − 50.0 Couple Rel