Technical data
380 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Fast Spurious Meas
Turns the fast spurious measurement test on or off. Pressing Fast Spurious Meas to On
provides a faster method of execution as with fast spurious testing spurs above the limit
line are reported. Any spurs reported outside the limit will cause the measurement to fail.
See “Abs Start Limit" on page 367 for more information.
Key Path: Meas Setup
Saved State: Ref Level state is saved to Instrument State. Ref Level value is recalculated
for each active range the value is not stored in Instrument State.
Factory Preset: Off
Range: Off|On
Remote Command:
[:SENSe]:SPURious:FSMeas OFF|ON|0|1
[:SENSe]:SPURious:FSMeas?
Example:
SPUR:FSM ON