Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 375
Remote Command: See “Detector 2" on page 371.
Off
Detector 2 is off, there is no second trace.
Key Path: Meas Setup, Range Table, Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
Points
Used to set the points mode of the N8201A. Setting to Man allows you to determine the
setting, or by selecting Auto the N8201A will determine the number of trace points to
ensure the bucket resolution equals RBW/2.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: Auto, the value is determined by the radio standard and the instrument
model.
Range: 101 to 8192
Remote Command:
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:POINts <number of
points>
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:POINts?
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:POINts:AUTO
1|0|ON|OFF
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:POINts:AUTO?