Technical data

374 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
Quasi Peak
Selects quasi peak detection. This is a fast-rise, slow-fall detector used in making CISPR
compliant EMI measurements. Quasi peak detection displays a weighted amplitude using
specific charge, discharge, and meter-movement time constants as described in CISPR
Publication 16.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
EMI Average
Selects EMI averaging detection. EMI Average detection displays the average value of
amplitude envelope. It is defined for EMI measurements by the CISPR standard and uses
the VBW filter as a low pass filter, the output of which is sampled during the bucket
interval. This functionality is quite different from the Average detector.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
EMI Peak
Selects peak detection using EMI specifications.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
MIL Peak
Selects peak detection using the MIL specifications.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.