Technical data

372 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Range: 0 to 70 dB
Remote Command:
[:SENSe]:SPURious[:RANGe][:LIST]:DETector[1]|2 [:FUNCtion]
OFF|NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak|EAVerage
|EPOSitive|MPositive
[:SENSe]:SPURious[:RANGe][:LIST]:DET2?
Example: SPUR:DET2 EAV
Normal
Displays the peak-detected level in the interval (bucket) being displayed when the signal is
CW-like. If the signal is noise-like (within a bucket the signal both rose and fell), the even
bucket shows the peak (maximum) within a two-bucket interval, and the odd bucket shows
the negative peak (minimum). Gain is increased to compensate for the effects of faster
sweep rates, to keep the displayed value of CW signals accurate.
When Normal is selected, Norm appears on the left side of the display.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.
Average
For each interval (bucket) in the trace, Average detection displays the average of all
samples within the interval using one of the following averaging methods:
log (also known as video)
power (also known as RMS)
voltage envelope
To explicitly set the averaging method, use the BW/Avg, Avg/VBW Type key. The
combination of the average detector and the power method is equivalent to what is
sometimes referred to as “RMS detection.” When the method (Avg/VBW Type) is set to
Pwr Avg, and Detector is set to Average, the RMS method is selected.
When Average is selected, Avg appears on the left side of the display.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in Instrument State.
Remote Command: See “Detector 2" on page 371.