Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 371
Range: 0 to 70 dB
Remote Command:
[:SENSe]:SPURious[:RANGe][:LIST]:DETector[1]|2 [:FUNCtion]
NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak|EAVerage|
EPOSitive|MPositive
[:SENSe]:SPURious[:RANGe][:LIST]:DET?
Example: SPUR:DET QPE
Detector 2
Allows you to set the detector to be used by the trace which will be used for display
purposes only.
The detector choices are:
Normal - displays the peak of CW-like signals and maximums and minimums of noise-like
signals.
Average - displays the average of the signal within the bucket. The averaging method
depends upon Avg Type selection (voltage, power or log scales).
Peak - displays the maximum of the signal within the bucket.
Sample - displays the instantaneous level of the signal at the center of the bucket
represented by each display point.
Negative Peak - displays the minimum of the signal within the bucket.
Quasi Peak - a fast-rise, slow-fall detector used in making CISPR compliant EMI
measurements.
EMI Average - displays the instantaneous level of the signal at the center of the bucket, just
like the sample detector. Also changes the auto coupling of VBW, RBW and Avg/VBW Type
and the set of available RBWs. Used in making CISPR-compliant measurements.
EMI Peak - the same as the Peak detector but uses CISPR related bandwidths.
MIL Peak - the same as the Peak detector but uses MIL related bandwidths.
Off - turns off second trace (Detector 2).
Key Path: Meas Setup, Range Table
State Saved: Saved in Instrument State.
Factory Preset: Auto, the value is determined by the radio standard and the instrument
model.