Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 369
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion <integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion?
Pk Threshold
Sets the minimum amplitude of signals that can be identified as peaks. For example, if a
value of –90 dBm is selected, only peaks that rise and fall more than the peak excursion
value and are above the Pk Threshold of –90 dBm are identified.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7 on page 363.
Range: –100 dB to 0.0 dB
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold <integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold?
Attenuation
Sets the attenuation for each range. This parameter can send up to 20 values. The location
in the list sent corresponds to the range the value is associated with. Missing values are
not permitted, for example, if you want to change values 2 and 6 you must send all values
up to 6. Subsequent values will remain as they were.
Key Path: Meas Setup, Range Table
State Saved: Saved in Instrument State.
Factory Preset: Auto, the value is determined by the radio standard and the instrument
model.