Technical data
368 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Dependencies/Couplings: If the limit is set to Auto, Abs Start Limit and Abs Stop Limit
are coupled to make a flat limit line. If set to Man, the absolute start and stop limits can take
different values to make a sloped limit line.
Factory Preset: See Table 7 on page 363.
Range: –150 dBm to 50 dBm
Remote Command:
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]
:DATA:STOP <integer>,<integer>
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]
:DATA:STOP?
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]
:DATA:STOP:AUTO OFF|ON|0|1, OFF|ON|0|1
:CALCulate:SPURious[:RANGe][:LIST]:LIMit:ABSolute[:UPPER]
:DATA:STOP:AUTO?
Remote Command Notes:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
The query for this command will always return 20 values.
Example:
:CALC:SPUR:LIM:ABS:DATA -23,-27
:CALC:SPUR:LIM:ABS:DATA:AUTO ON, ON, ON, ON
Peak Excursn
Sets the minimum amplitude variation of signals that can be identified as peaks. For
example, if a value of 6 dB is selected, peaks that rise and fall more than the 6 dB above the
peak threshold value are identified.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7 on page 363.
Range: 0.0 dB to 100.0 dB
Remote Command: