Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 365
Res BW
Sets the resolution bandwidth of the N8201A. When Auto is selected the N8201A
determines the optimum setting, while Man enables you to determine the setting.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7 on page 363.
Range: Resolution bandwidth range of your N8201A.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]
:AUTO OFF|ON|0|1 selects the mode.
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]<
integer>
The query for this parameter always returns 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]
:AUTO?
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]?
Video BW
Sets the video bandwidth of the N8201A. When Auto is selected the N8201A determines
the optimum setting, while Man enables you to determine the setting.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7 on page 363.
Range: Video bandwidth range of your N8201A.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.