Technical data
364 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Start Freq
Used to set the start frequency of the N8201A.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7 on page 363.
Range: Frequency range of your N8201A.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt <integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt?
Stop Freq
Stop Freq is used to set the stop frequency of the N8201A.
Key Path: Meas Setup, Range Table
Factory Preset: See “Range Table” on page 204.
Range: Frequency range of your N8201A.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP
<integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STOP?