Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 363
Points = (Man) 601
The table below defines the other parameter’s settings for the measurement. The values for
ranges 1 through 5 are based on the W-CDMA (3GPP) BTS Spurious Emission
measurement.
Remote Command: There is no equivalent remote command.
Range
Selects a range and updates the values on the other Range Table keys so that they reflect
the settings for the selected range. If Range is set to On it is used as part of the
measurement; when set to Off it is excluded.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 7.
Range: Off|On
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted. For example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values remain
as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe?
Example: SPUR:STAT OFF,ON,ON,OFF,ON,ON lists values 1 through 6
Table 7 Range Table Default Settings
Range Start Freq (GHz) Stop Freq (GHz) Res BW (kHz)
1 1.920 1.980 1200
2 1.894 1.920 510
3 2.100 2.102 100
4 2.175 2.180 100
5 0.800 1.000 4000
6-20 1.5 2.5 Auto