Technical data

Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 361
Avg Mode
Selects the type of termination control used for the averaging function (Exp or Repeat). This
determines the averaging action after the specified number of measurements (average
count) is reached.
EXP (Exponential Averaging mode) When you set Avg Mode to Exp, each successive
data acquisition after the average count is reached is exponentially weighted and combined
with the existing average. Exponential averaging weights new data more than old data,
which facilitates tracking of slow-changing signals. The average will be displayed at the
end of each sweep.
Repeat When you set Avg Mode to Repeat, after reaching the average count, all previous
result data is cleared and the average count is set back to 1.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: EXPonential
Range: EXPonential|REPeat
Remote Command:
[:SENSe]:SPURious:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:SPURious:AVERage:TCONrol?
Example:
SPUR:AVER:TCON EXP
SPUR:AVER:TCON?