Technical data

358 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Optimize Ref Level
Sets the reference level to place the tones near the top of the display. When input
attenuation is set to Auto, pressing Optimize Ref Level sets the input attenuation according
to the auto rules described in Max Mixer Lvl, based on this reference level. You may want to
trade off N8201A noise and N8201A-induced distortion by manually changing the
attenuation setting from this starting point.
There is always a minimum of 2 dB of attenuation set to protect the N8201A input.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example:
POW:RANG:AUTO ONCE
POW:RANG:AUTO?