Technical data

340 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Center Freq
Used to set the center frequency of the N8201A.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 6.
Range: Frequency range of your N8201A.
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated. Missing values are not permitted, for example, if you want to
change values 2 and 6, you must send all values up to 6. Subsequent values will remain as
they were.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency <integer>
The query for this parameter will always return 10 values.
[:SENSe]:HARMonics:RANGe[:LIST]:FREQuency?
Example: HARM:RANG:LIST:FREQ 300
Span
Span is used to set the span of the N8201A.
Key Path: Meas Setup, Range Table
Factory Preset: See Table 6.
Range: Span of your N8201A.
Remote Command:
This parameter can send up to 10 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN
[:SENSe]:HARMonics:RANGe[:LIST]:SPAN?