Technical data

330 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Meas Interval
Specifies the time interval over which the measurement is made. The minimum setting is
100 ms. The maximum setting depends on the measurement bandwidth setting but never
exceeds 600 ms. The resolution is 1 ms.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: 1.0ms
Default Terminator: s
Range: 100 ms to 600 ms
Remote Command:
[:SENSe]:PSTatistic:SWEep:TIME <time>
[:SENSe]:PSTatistic:SWEep:TIME?
Example:
PST:SWE:TIME 1.0e–3
PST:SWE:TIME?