Technical data
318 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Default Terminator: dB
Range: –200 dB to 200 dB
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]
:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]
:DATA?
Example:
:CALC:MCP:OFFS:LIST:LIM:POS:DATA -50dB,-55dB,-60dB
:CALC:MCP:OFFS:LIST:LIM:POS:DATA?
Lower Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the left of the carriers.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: 0 dB, 0 dB, 0 dB
Default Terminator: dB
Range: –200 dB through 200 dB
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]
:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]
:DATA?
Example:
CALC:MCP:OFFS:LIST:LIM:NEG:DATA -50dB,-55dB,-60dB
CALC:MCP:OFFS:LIST:LIM:NEG:DATA?