Technical data

Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 317
Example:
MCP:OFFS:LIST 5000HZ,7.5000HZ,15000HZ
MCP:OFFS:LIST?
Offset Integ BW
Sets the bandwidth over which the power is integrated for the selected offset (refer to
“Offset” above).
Key Path: Meas Setup, Offset/Limits
Key Notes: If RRC Filter is set to On, the actual integration bandwidth used is the
displayed integration bandwidth multiplied by (1 + Filter Alpha).
State Saved: Saved in Instrument State.
Factory Preset: 3.84 MHz, 3.84 MHz, 3.84 MHz
Te rm in a tor s: Hz, kHz, MHz, GHz
Default Terminator: Hz
Range: 100 Hz to 20 MHz
Remote Command:
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:
[INTegration]<Hz>,<Hz>,<Hz>
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:
[INTegration]<Hz>,<Hz>,<Hz>
Example:
MCP:OFFS:LIST:BWIDth 5MHz,3MHz,5MHz
MCP:OFFS:LIST:BWID?
Upper Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the right of the carriers.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: 0 dB, 0 dB, 0 dB