Technical data

314 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Example:
MCP:CARR:LIST:PPR YES,NO,YES,NO,YES
MCP:CARR:LIST:PPR?
Carrier Width
Specifies the width of each carrier including carriers with no power present. There is a
corresponding one-to-one relationship between each carrier width value and the specified
number of carriers.
1 Press the Carrier key and select the carrier number you wish to define. (The carrier
number selected is shown on the Carrier key.)
2 Press the Carrier Width key and enter the width using the keyboard.
Key Path: Meas Setup, Carrier Setup, Configure Carriers
Key Notes: The Carrier key determines which carrier width you are defining.
State Saved: Saved in Instrument State.
Factory Preset: 5 MHz, 5 MHz, 5 MHz, 5 MHz
Te rm in a tor s: Hz, kHz, MHz, GHz
Default Terminator: Hz
Resolution: 1 Hz
Range: 0 Hz to 500 MHz
Remote Command:
:SENSe]:MCPower:CARRier:LIST:WIDTh <Hz>
[:SENSe]:MCPower:CARRier:LIST:WIDTh?
Remote Command Notes:
The position number of the each carrier width in the list sent corresponds to the carrier
number you are defining. For example: “5 MHz, 10 MHz, 5 MHz” defines six carriers.
Carriers 1, 2, 4, and 5 are defined as having power present whereas carriers 3 and 6 do not
have power present. If you need to change carrier 5, you must send all carriers up to 5.
Carrier 6 will remain unchanged. If you send more values than the number of carriers
specified using the Carriers key, (in this example, there are six carriers) they will be ignored.
The query for this parameter returns the current value for all carriers (with and without
power present).