Technical data

306 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Noise Correction
Turns noise correction on or off. When you set Noise Corr to On, a calibration of the noise
floor is performed and used to correct for N8201A noise floor contribution to measurement
levels, increasing dynamic range.
Noise Correction is unavailable if signal tracking is on, or when the radio standard is set to
NADC or PDC when the measurement method is set to RBW.
Key Path: Meas Setup,
State Saved: Saved in Instrument State.
Factory Preset: Off
Remote Command:
[:SENSe]:ACPower:CORRection:NOISe[:AUTO] OFF|ON|0|1
[:SENSe]:ACPower:CORRection:NOISe[:AUTO]?
Remote Command Notes:
The noise correction feature is not available when the radio standard is set to NADC or PDC
Example:
ACP:CORR:NOIS 1
ACP:CORR:NOIS?