Technical data

Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 301
Total P wr R e f
Enables you to set the adjacent channel power reference to automatic or manual when
Meas Type is set to Total Pwr Ref. When set to automatic, the carrier power result reflects
the measured power value in the carrier. When set to manual, the last measured value is
captured and held, or may be entered by the user. Relative values are displayed, referenced
to the total power measured in the main channel.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: Auto, Measured carrier power value
Te rm in a tor s: dBm
Default Terminator: dBm
Range: –200 dBm to 200 dBm
Remote Command:
[:SENSe]:ACPower:CARRier[:POWer]
[:SENSe]:ACPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:CARRier:AUTO[:STATe]?
Example:
ACP:CARR:AUTO 0
ACP:CARR:AUTO?
PSD Ref
This function is only available when measurement type is set to PSD Ref.
Enables you to set the power spectral density in the carrier (main channel) that will be used
to compute the relative power spectral density values for the offsets when Meas Type is set
to PSD Ref. When the PSD Ref state is set to Auto, this will be set to the measured carrier
power spectral density.
Factory Preset: 0 dBm
Te rm in a tor s: dBm
Default Terminator: dBm