Technical data
Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 299
Fast
Sets the measurement control method to Fast. When Fast method is selected, the 5 point
ACLR measurement method is used. The following parameters are configured to allow the
measurement to be made. This method is only available for the 3GPP W-CDMA radio
standard, this key will be disabled for all other radio standards.
The current function values are not saved when fast method is turned on. Most functions
are disabled when Fast is selected.
Function Value
Channel Integration BW 3.84 MHz
Offset State On, On, Off, Off, Off, Off
Offset Frequency 5 MHz, 10 MHz, 0 Hz, 0Hz, 0 Hz
Ref BW 3.84 MHz
Meas Type Total Pwr Ref
Opt Ref Lvl Disabled
RRC Filter State On
Filter Alpha 0.22 (T fixed at 1/3.84 MHz)
Noise Correction Off
Radio Std
*
3 GPP W-CDMA
Sweep Time
*
,
†
6.2 ms
Sweep Time Mode
*
Manual
Gate Off
Sweep Points
*
5
Trigger Free Run
Detector Average
Res BW 470 kHz
Res BW Mode Manual
Video BW 5 MHz
Video BW Mode Auto
Avg/VBW Type Power
Avg/VBW Type Mode Auto
Span 23.84 MHz
Span Zoom Not Applicable