Technical data

298 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Method
Enables you to set the measurement method. The resolution bandwidth method is most
useful for measuring cdmaOne and cdma2000 signals; the integration bandwidth method is
preferred for other signals.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: IBW
Remote Command:
[:SENSe]:ACPower:METHod IBW|RBW|FAST
[:SENSe]:ACPower:METHod?
Example:
ACP:METH RBW
ACP:METH?
IBW
Sets the measurement control method to IBW.
Key Path: Meas Setup, Method
Remote Command: Use[:SENSe]:ACPower:METHod. See “Method” on page 91.
Example: ACP:METH IBW
RBW
Sets the measurement control method to RBW.
Key Path: Meas Setup, Method
Remote Command: Use[:SENSe]:ACPower:METHod. See “Method” on page
91.
Example: ACP:METH RBW