Technical data

296 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Meas Type
Specifies the reference for the measurement, either Total Pwr Ref or PSD Ref. Relative
values can be displayed referenced to either the total power (Total Pwr Ref) or the power
spectral density (PSD Ref) measured in the main channel.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: Total Pwr Ref
Remote Command:
[:SENSe]:ACPower:TYPE PSDRef|TPRef
[:SENSe]:ACPower:TYPE?
Example:
ACP:TYPE PSDR
ACP:TYPE?
Total Pw r Ref
When Total Pwr Ref is selected, the power in the carrier is used as the reference in
computing the relative power values for the offsets.
Key Path: Meas Setup, Meas Type
PSD Ref
When PSD Ref is selected, the mean power spectral density of the carrier is used in
computing the relative power spectral density values for the offsets.
Key Path: Meas Setup, Meas Type
Factory Preset: On (Auto)