Technical data
294 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Example:
ACP:OFFS:LIST:BAND 50,50,50,50,50,50
ACP:OFFS:LIST:BAND?
Pos Offset Limit
Enables you to set the upper limit for the upper segment of the specified offset pair.
Key Path: Meas Setup, Offset Setup
State Saved: Saved in Instrument State.
Factory Preset: 0.00 dB
Default Terminator: dB
Range: –200 dB to 200 dB
Remote Command:
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA
<rel_power>,<rel_power>,<rel_power>,<rel_power>,
<rel_power>,<rel_power>
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]
:DATA?
Example:
CALC:ACP:OFFS:LIST:LIM:POS:DATA 10,10,10,10,10,10
CALC:ACP:OFFS:LIST:LIM:POS:DATA?
Neg Offset Limit
Enables you to set the upper limit for the lower segment of the specified offset pair.
B OFF 0 Hz 2 MHz
C OFF 0 Hz 2 MHz
D OFF 0 Hz 2 MHz
E OFF 0 Hz 2 MHz
F OFF 0 Hz 2 MHz
Offset State Frequency Integ. BW