Technical data

282 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Occ BW % Pwr
Enables you to change the percentage of signal power used when determining the occupied
bandwidth.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: 99.0%
Range: 10.0% through 99.99%
Remote Command:
[:SENSe]:OBWidth:PERCent <percent>
[:SENSe]:OBWidth:PERCent?
Example:
OBW:PERC 98
OBW:PERC?