Technical data

Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 273
Integ BW
Specifies the range of integration used in calculating the power in the channel. Note that
the integration bandwidth is displayed on the trace as two markers connected by an arrow.
Be sure the Span of the instrument is set between 1 and 10 times the integration
bandwidth.
Key Path: Meas Setup
Key Notes: If RRC Filter is on, the actual integration bandwidth used is the displayed
integration bandwidth multiplied by (1 + “Filter Alpha”).
State Saved: Saved in Instrument State.
Factory Preset: 2 MHz, or as defined by the selected radio standard.
Te rm in a tor s: GHz, MHz, kHz, Hz
Default Terminator: Hz
Range: 100 Hz to maximum span.
Remote Command:
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Example:
CHP:BAND:INT 1 MHz
CHP:BAND:INT?