Technical data
266 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
14 Measure (Spectrum Analysis Mode)
n=9 Returns 12 comma-separated scalar values of frequency (in
Hz) that have peak power in each offset (14 when offset F is
active):
1. Reserved for the future use, returns –999.0
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11.Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
n=10 Returns 12 comma-separated scalar values (in dBm) of the
absolute peak power of the segment frequencies (14 when
offset F is active):
1. Reserved for the future use, returns –999.0
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11.Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
n=11 Returns 12 comma-separated scalar values (in dBc) of the
peak power relative to the carrier at the segment
frequencies (14 when offset F is active):
1. Reserved for the future use, returns –999.0
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11.Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
n=12 Returns the band power result when in WLAN standard. If
not in WLAN standard the value returned will be –999.0
n=13 Returns the number of failed trace points
Measurement Results Available (continued)
Command / Condition n Results Returned