Technical data
264 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
14 Measure (Spectrum Analysis Mode)
Using Total Power
Reference
n=5 Total Power Reference
Returns 12 comma-separated scalar values (in dBm) of the
absolute power of the segment frequencies (14 when offset
F is active):
1. Total power reference (dBm)
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11.Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
Using Power Spectral
Density Reference
n=5 Power Spectral Density Reference
Returns 12 comma-separated scalar values (in dBm/Hz) of
the absolute power of the segment frequencies (14 when
the offset F is active):
1. Total power reference (dBm)
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11. Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
Using Total Power
Reference
n=6 Total Power Reference
Returns 12 comma-separated scalar values (in dBc) of the
power relative to the carrier at the segment frequencies (14
when offset F is active):
1. Reserved for the future use, returns –999.0
2. Reserved for the future use, returns –999.0
3. Negative offset frequency (A)
4. Positive offset frequency (A)
…
11.Negative offset frequency (E)
12.Positive offset frequency (E)
If offset F is active:
13.Negative offset frequency (F)
14.Positive offset frequency (F)
Measurement Results Available (continued)
Command / Condition n Results Returned