Technical data

258 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
14 Measure (Spectrum Analysis Mode)
Key Path: MEASURE
Remote Command:
Example:
FETC:BPOW? or MEAS:BPOW? or READ:BPOW?
Measurement Results Available
Command n Results Returned
:CONFigure:BPOWer
:INITiate:BPOWer
N/A Not Applicable
:FETCh:BPOWer[n]?
:MEASure:BPOWer[n]?
:READ:BPOWer[n]?
n=1 (or not
specified)
Returns 10 scalar results, in the following order.
Sample time
Power (RMS of carrier power)
Power averaged
Number of samples,
Amplitude threshold (relative)
Maximum trace point on the measured part of the burst.
Minimum trace point on the measured part of the burst.
Burst width (actual measured width, using the threshold
level)
Measured time
Measured points
n=2 Returns the displayed RF envelope trace data separated
by commas. The number of data values is determined by
the setting of the trace points parameter.