Technical data

Measure (Spectrum Analysis Mode) 14
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 257
where Pavg = average power, n is the start trace point, m= the stop trace point, and p= the
trace point amplitude power in dBm.
Figure 7 shows an example of the results obtained when measuring a Bluetooth™ signal
and with a user-defined burst width.
Figure 7 Burst Power Measurement Results
The analyzer defaults to zero-span mode and the sweep time is set to capture at least one
burst. The sweep time can be changed by pressing Sweep, Sweep Time.
Pressing Meas Setup after Burst Power has been selected displays the burst power
measurement setup menu. Pressing Meas Control after Burst Power has been selected
displays the burst power control menu, where you can pause or restart a measurement, or
toggle between continuous and single measurement.
Key Access: MEASURE
The measurements described above are those available in SA mode (see Mode key). Other
measurements are available in other modes if an optional personality is installed.
Pavg()10 10
10
p
10
-----
⎝⎠
⎜⎟
⎛⎞
n
m
⎝⎠
⎜⎟
⎜⎟
⎛⎞
mn
-------------------------------------
⎩⎭
⎪⎪
⎪⎪
⎪⎪
⎨⎬
⎪⎪
⎪⎪
⎪⎪
⎧⎫
log=