Technical data
Measure (Spectrum Analysis Mode) 14
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 245
not
specified,
or n=1
Meas Type = Total
Power Reference
Method = IBW or
RBW
Returns 28 comma-separated values, in the
following order:
Center Frequency - relative power (dB)
Center Frequency - absolute power (dBm)
Center Frequency - relative power (dB)
Center Frequency - absolute power (dBm)
Negative Offset Frequency (1)
- relative power (dB)
Negative Offset Frequency (1)
- absolute power (dBm)
Positive Offset Frequency (1)
- relative power (dB)
Positive Offset Frequency (1)
- absolute power (dBm)
...
25. Negative Offset Frequency (6)
- relative power (dB)
26. Negative Offset Frequency (6)
- absolute power (dBm)
27. Positive Offset Frequency (6)
- relative power (dB)
28. Positive Offset Frequency (6)
- absolute power (dBm)
The center frequency relative power is relative to
the center frequency absolute power, and
therefore is always equal to 0.00 dB.
not
specified,
or n=1
Meas Type = Total
Power Reference
Method = Fast
Returns 5 comma-separated values, in the
following order:
Main Carrier- absolute power (dBm)
Negative Offset Frequency (1)
- absolute power (dBm)
Positive Offset Frequency (1)
- absolute power (dBm)
Negative Offset Frequency (2)
- absolute power (dBm)
Positive Offset Frequency (2)
- absolute power (dBm)
Measurement Results Available (continued)
Command n Condition Results Returned