Technical data

Measure (Spectrum Analysis Mode) 14
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 241
Remote Command:
Remote Command Notes:
The main channel power is returned in the current amplitude units, and the density value is
returned in current amplitude units/Hz
Example:
FETC:CHP? or MEAS:CHP? or READ:CHP? command returns scalar results of main
channel power and power density.
FETC:CHP:CHP? or MEAS:CHP:DENS? or READ:CHP:DENS? commands will
return the single scalar result specified.
Measurement Results Available
Command Results Returned
:CONFigure:CHPower
:INITiate:CHPower
Not Applicable
:FETCh:CHPower?
:MEASure:CHPower?
:READ:CHPower?
Returns two values that correspond to the Channel Power and
Power Spectral Density.
:FETCh:CHPower:CHPower?
:MEASure:CHPower:CHPower?
:READ:CHPower:CHPower?
Returns a single value that corresponds to the Channel Power.
:FETCh:CHPower:DENSity?
:MEASure:CHPower:DENSity?
:READ:CHPower:DENSity?
Returns a single value that corresponds to the Power Spectral
Density.