Technical data

Meas Control 13
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 229
Measure
Switches the N8201A between triggering the current measurement/sweep continuously or
triggering a single measurement. The front panel Single key also puts the N8201A in
single-measurement mode.
Key Path: Meas Control
State Saved: Save
Factory Preset: Continuous
Remote Command:
Use :INITiate:CONTinuous OFF|ON. See “" on page 459.
Remote Command Notes:
This command affects sweeping when in the SA mode. It affects measurements when a
measurement has been selected from the MEASure command subsystem.
When ON, at the completion of each trigger cycle, the trigger system immediately
initiates another trigger cycle.
When OFF, the trigger system remains in an “idle” state until CONTinuous is set to ON
or an :INITiate[:IMMediate] command is received. On receiving the
:INITiate[:IMMediate] command, it will go through a single trigger cycle, and then return
to the “idle” state.
The query INIT:CONT? returns 1 or 0. 1 is returned when the instrument is continuous
triggering. 0 is returned when it is single triggering.
Example:
INIT:CONT OFF