Technical data
222 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
12 Marker Fctn
Remote Command:
See “Marker Fctn" on page 219 for the command to select a function.
Remote Command Notes:
Note that the value when the Y-axis units are watts is the square of the value when the
Y-axis units are volts. For example, when the percent ratio with Y-axis units in volts is 20%
(0.2), the percent ratio with Y-axis units in watts will be 4% (0.2
2
= 0.04). When you read the
value out remotely you have to know whether you are in log (dB) or linear (percent), and if
linear, whether volts or watts.
Example:
CALC:MARK:FUNC NOIS turns on marker 1 as a noise marker.
CALC:MARK:FUNC? returns the current setting of marker function for the marker
specified. In this case it returns the string: NOIS.
CALC:MARK:Y? returns the y-axis value of the Marker Noise function for marker 1 (if
Marker Noise is ON for marker 1).